15"/12"/10.4" high-resolution display touch screen. Comprehensive calculations for clinical application.
Flat design: Clear and concise menu, no difficulty for doctors and nurses to learn, and any routine operation can be completed in 2 steps at most, significantly improving the efficiency of use.
Simultaneous 12-lead ECG: Support 3/5/6/10 multiple electrode connections, up to 12 leads of ECG can be displayed on the same screen, which is equivalent to equipping a professional ECG machine.
Portable and easy to move.
Powerful clinical features.
Minimalist Design:
* Capacitive touch screen, easy and convenient to operate.
* Ultra-high resolution provides clearer picture quality and better visual effects.
* Tilted screen ergonomically designed for better bedside viewing.
* Dual alarm lights facilitate long-distance identification of alarm conditions.
* Fanless design reduces noise to provide a better monitoring environment.
Excellent IT workflow.
Technical Specification | |
ECG | |
Input dynamic range | ±(0.5mVp~5mVp) |
Differential input impedance | ≥10MΩ |
Bandwidth | 0.05~150Hz (Diagnostic) 0.5~40Hz (Monitoring) 1~20Hz (Operation) |
CMRR | ≥90dB (Diagnostic) ≥105dB (Monitoring & Operation) |
Sensitivity selection | ×1/4, ×1/2, ×1, ×2, ×4 and Auto |
Sweeping speed | 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s |
HR measuring range | 15~350bpm |
HR accuracy | ±1% or ±2bpm, whichever is greate |
Pacemaker pulse detection and rejection function | |
RESP | |
Measuring range | 0~120rpm |
Measuring accuracy | ±5% or ±2 rpm, whichever is greater |
TEMP | |
Measuring range | 0~50℃ |
Measuring accuracy | ±0.1℃ |
NIBP | |
Technique | Oscillometric method |
Typical measurement time | <30 seconds (adult cuff) |
NIBP measuring range | SYS: 40~275mmHg (Adult) 40~200mmHg (Pediatric) 40~135mmHg (Neonate) |
NIBP measuring range | DIA: 10~210mmHg (Adult) 10~150mmHg (Pediatric) 10~95mmHg (Neonate) |
NIBP measuring range | MAP: 20~230mmHg (Adult) 20~165mmHg (Pediatric) 20~110mmHg (Neonate) |
NIBP measuring accuracy | Mean difference: ±5mmHg Standard deviation: 8mmHg |
NIBP measurement mode | Manual, Auto, STAT, Multi-cycle mode |
Auto measuring intervals | 1-480min |
SpO2 | |
Technique | Dual-wavelength optical method |
Measuring range | 0%~100% |
Measuring accuracy | Arms is not greater than 2% for SpO2 range 70~100%. |
PR measuring range | 30~250bpm |
PR measuring accuracy | ±2bpm or ±2%, whichever is greater |
Low perfusion performance | As low as 0.3%. |
CO2 | |
Technique | Infrared optical method |
Sampling mode | Sidestream or Mainstream |
Measuring range | 0~150mmHg |
Measuring accuracy | 0~40mmHg, ±2mmHg 41~70mmHg, ±5% of reading 71~100mmHg, ±8% of reading 101~150mmHg, ±10% of reading |
Flow rate | 50ml/min ±10 ml/min (Sidestream) |
Cerebral State Monitoring (CSM) | |
EEG sensitivity | ±400µV |
Noise level | <2µVp-p, <0.4µV rms (1~250Hz) |
CMRR | >140dB |
Input impedance | >50Mohm |
CSI and update | 0-100. filter: 6-42Hz, 1 sec. update |
EMG% | 0-100 (logarithmic) filter: 75-85 Hz, 1 sec. update |
BS% | 0-100. filter: 2-42 Hz, 1 sec. update |
IBP | |
Technique | Strain gauge transducer |
Input sensitivity | 5µV/V/mmHg |
Measuring range | -50~300mmHg |
Measuring accuracy | ±2% or ±4mmHg, whichever is greater |
Measuring positions | ART, RAP, PA, LAP, CVP, ICP, AUXP1, AUXP2 |
Calibration | zero calibrating |
Cardiac Output (C.O.) | |
Blood temperature measuring range | 23-43℃, accuracy: ±0.5℃ |
Injecta temperature measuring range | 0-20℃, accuracy: ±0.5℃ |
Measuring range | 0.2~20 L/min |
Measuring accuracy | ±0.2 L/min or ±10%, whichever is greater |
Other Specifications | |
Power supply | AC 100V-240V, 50/60Hz, 60VA |
Built-in lithium battery | 11.1V/4400mAh |
Display | 15 inch TFT display |
Alarming method | 3 levels audible-visible alarm |
Networking | Ethernet |
Standard configuration | |
ECG, Respiration, SpO2, PR, NIBP, Temperature | |
Options | |
Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG, Cardiac Output, Cerebral State Monitoring, Central Monitor Station, Multi-Gas Monitoring, Wifi |
Thank you for your attention on LEPU
Related Lepu-Creative Multiparameter Patient Monitors