Principle Of Pulse Oximeter
2007-6-5
 

The principle of Pulse Oximeter is based on the red and infrared light absorption characteristics of oxygenated and deoxygenated hemoglobin. Oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Deoxygenated (or reduced) hemoglobin absorbs more red light and allows more infrared light to pass through. Red light is in the 600-750 nm wavelength light band. Infrared light is in the 850-1000 nm wavelength light band. 

Pulse Oximeter uses a light emitter with red and infrared LEDs that shines through a reasonably translucent site with good blood flow. Typical adult/pediatric sites are the finger, toe, pinna (top) or lobe of the ear. Infant sites are the foot or palm of the hand and the big toe or thumb. Opposite the emitter is a photodetector that receives the light that passes through the measuring site

Remark:  Our company provides various pulse oximeter, such as  Finger Oximeter, Handheld Pulse Oximeter and Wrist Oximeter, etc.. What is more, we also provide Pulse Oximeter Module and OEM service.