KRK NIBP Module

Model:12020320

The NIBP Module combines the hardware and software of Non-Invasive measurement of Blood Pressure. The NIBP Module has the following features:
Automatic oscillometric technology.
Appropriate to adult, pediatric, and neonate
4 working modes with Automatic, Manual, STAT and Calibration
Step deflation with artifact suppression algorithm
Low power consumption.

Specifications

Size: 90mm(L) x 70mm(W) x 38mm (H)
Weight: 

250g

Power supply:
12VDC
Power consumption: max. current during measurement: 200mA;
max. current during inflation: 300mA
max. current during stand: 50mA
Communications: UART interface with TTL level, default baud rate: 4800bps

Measurement Ranges:

Adult:
Pediatric:
Neonate:
Systolic: 40 to 270 mmHg
MAP: 20 to 230 mmHg
Diastolic: 10 to 210 mmHg
Pulse Rate: 40 to 240 bpm
Systolic: 40 to 200 mmHg
MAP: 20 to 165 mmHg
Diastolic: 10 to 150 mmHg
Pulse Rate: 40 to 240 bpm
Systolic: 40 to 135 mmHg
MAP: 20 to 105 mmHg
Diastolic: 10 to 95 mmHg
Pulse Rate: 40 to 240 bpm

Resolution
Pressure: 1 mmHg
Pulse Rate: 1 bpm

Accuracy
NIBP: Meets ANSI/AAMI SP10-1992
Pulse Rate: ±2% or ±2 bpm, whichever is greater
Pressure measuring accuracy: ±3mmHg